News
Kaspersky Endpoint Security interferes with F3 camera operation.
Posted by Stephen Barnard on 14 May 2019 09:57 am

Kaspersky Endpoint Security 10/11 may cause camera connection issues with the Synoptics FlexCam models of camera that are used in the Syngene GBox F3, InGenius3 and UGenius3.

GeneSys will display the "connecting to camera" icon and stay that way indefinately.

Kaspersky has a "Block access to video capturing devices" function which can block these cameras.

To disable this function :

Open the Kaspersky Security Dialog by clicking on the "K" icon in the Windows Notification area of the taskbar.

In KES 10, select the Settings tab.  In KES 11, select "Protection Components", then "Application Control".

In KES 10, Select "Endpoint Control", then "Application Privilege Control", then "Applications".  In KES 11, Select "Essential Threat Protection", then select "Firewall", then "Application Rules".

In either package, scroll down to find "Low Restricted".  Right click on "Low Restricted" and select "Group rules".

Select the "Rights" tab.  Scroll down to "Suspicious system modifications" and set "Access to video capturing devices" to "allow".

Press "ok" to return to the main screen and press save.

Restart your image capture software and the camera should function as normal.

We have no reports of other AntiVirus packages causing any problems, but they may have similar features.


Comments (0)
Post a new comment
 
 
Full Name:
Email:
Comments:
CAPTCHA Verification 
 
Please enter the text you see in the image into the textbox below (we use this to prevent automated submissions).